| Optical System |
| Principle |
Absorption spectrometry |
| Range of measurement |
From -0.1 to 3.0A |
| Normal Spectral range |
Automatic via filter wheel |
| Filters |
340, 405, 505, 546, 578, 630, 670 nm. 1 extra position |
| Bandwidth |
10nm (±2 nm) |
| Reading error |
2% at 2.5 A |
| Light source |
Halogen lamp |
| Zero |
Automatic |
| Settings |
Monochromatic & Bichromatic readings. |
| Detector |
Schottkey type photodiode - Silicon |
|
| Thermostat System |
| Peltier regulated |
25°, 30° & 37°C |
| Programmed Accuracy |
0.2°C |
|
| Cuvettes - Dual Mode |
| Flow cell of |
18 µl |
| Removable cuvettes |
Macro, Semi-micro & Micro |
|
| Printer |
| Type |
Fast Thermal Graphic |
| Printing speed |
50 mm/sec |
| Dots per line |
384 dots/line |
| Paper |
Thermo sensitive, 57 mm wide |
|
| Screen and Keyboard |
| LCD blue LED back lit |
240 x 64 pixels. |
| Keyboard |
Tactile Soft-touch |
|
| Aspiration System |
| Continuous flow system with Bi-directional peristaltic pump |
|
| Incubator |
| 37°C for 9 Position Test Tubes & 9 Position Cuvettes |
|
| Programming |
| Table of tests |
up to 200 |
| Table of units |
up to 9 units |
| Quality control storage of the last 30 results |
| Memory |
128 KB EEPROM |
|
| Quality Control |
| 2 Controls per test - Normal & Abnormal |
|
| Power / Dimensions |
| Voltage |
110-220 V |
| Frequency |
50 - 60 Hz |
| Maximum power |
80W |
| Temperature |
10° to 40° C |
| Maximum relative humidity |
80% |
| Dimensions |
40 cm x 32 cm x 16 |
|