• Biochemistry Analyzer

    Biochemistry Analyser / Elisa Reader / Coagulation

  • Bilirubin Analyzer

  • HBA1C Analyzer

  • Biochemistry Analyzer

  • Biochemistry Analyzer

Bio Chemistry Analyser

ARX-100

  • Main Feature
  • Technical Specification

SALIENT FEATURES
• User friendly software
• Editing of patient ID, Name (12 characters), Factor
• Print out of laboratory name & Doctor's name.
• For Kinetic test : On line Graph & OD. Graph editing facility. Estimation of result.
• Universal input voltage 90 to 265 V/50 - 60 Hz.
• Storage of 1000 test results in the memory with patient ID.
• Lamp Saver mode.
• Auto Filter Seletion.

METHODS OF ASSAY
• ABS
• EP with Factor
• KIN
• Fixed time
• Differential
• Multipoint upto 10 standards

Mixing by magnetic motor / stirrer INR / ISI Value % / Seconds Auto / Manual start
Parameters PT / APTT / Fibrinogen
Filter 405 nm / 630 nm.
Wells 8 position
Filters 405 nm/450 nm / 630 nm
Methods Cut-off / Multi standard with EP / KIN / Fix Time
O.D. Range 0 - 3.5A
Principle Absorption spectrometry
Range of measurement From – 0.1 to 3.0A
Nominal spectral range Automatic via filter wheel
Filters 340, 405, 450, 505, 546, 578, 630 and 670nm interference type
Bandwidth 10 nm (± 2nm)
Reading error ± 5% at 0.10 A
± 2% at 1.00 A
± 2% at 2.00 A
Light source Halogen lamp
Zero Automatic
Settings Monochromatic and Bichromatic readings.
Detector Schottkey type Photodiode - Silicon
Peltier regulated 25°, 30° & 37°C programmed
Accuracy ± 0.2°C
Flow cell of 18 µL
Removable cuvettes Macro, Semi-micro & Micro
Type Fast Thermal Graphic
Printing speed 50 mm/sec
Dots per line 384 dots/line
Paper Thermosensitive - 57 mm wide, Easy Loading
Programming On/Off test wise
Screen Graphic LCD 240 x 128 pixels
Back Light Blue LED Back lit
Keyboard Tactile Soft - Touch
Continuous flow system with peristaltic gear pump.
Sipping volume can be programmed from 200µ l to 5.0 ml.
Table of tests up to 235
Table of units up to 9 units
Quality control storage of the last 30 results
Memory 512 KB EEPROM
2 Controls per test - Normal & Abnormal
Levy-Jennings charts
Westgard's rules
Voltage 110-220 V
Frequency 50 - 60 Hz
Maximum power 110W
Temperature 10° to 40° C
Maximum relative humidity 80%
Dimensions 45 cm x 41 cm X 17
Weight 8 kgs