• Biochemistry Analyzer

    Biochemistry Analyser / Elisa Reader / Coagulation

  • Bilirubin Analyzer

  • HBA1C Analyzer

  • Biochemistry Analyzer

  • Biochemistry Analyzer

Elisa Plate Reader

  • Main Feature
  • Technical Specification

SALIENT FEATURES
• Multiple Plate Formats
• 12 / 24 / 48 and 96 well capability to meet a variety of assay needs
• Ensures optimal performance provides reliable, long term use
• On-Board Diagnostic Testing
• Storage of 1000 test results in the memory with patient ID.
• System Test checks the entire optics path with12 channel optics


METHODS OF ASSAY
• Absorbence
• End point
• Kinetic
• CUTOFF
• Multipoint

Absorbance / End point / Kinetic / CUTOFF and Multi point
Light source Tungsten halogen
Filter wheel 6 positions (2 optional)
Wavelength range 400 - 670 nm
Filters 405, 450, 492 and 630nm
Measurements From 0.0 to 3.5A
Bandwidth 10nm
Dynamic range 0 to 3.5 OD
Resolution 0.001 OD
Monochromatic Reading time 96 wells 12 seconds
Bichromatic readings Reading time 20 seconds
OD accuracy <1% at 2.5 OD
OD linearity <1% at 2.5 OD
<3% at 3.0 OD
OD repeatability <1% at 2.5
Screen Graphic LCD 240x128 pixels
Keyboard Tactile Soft - Touch
Printer Built-in 57mm thermal printer
External Printer Optional
PC connection Optional
100 user-programmable Tests
96-well Reader Control Software and Data Analysis for raw data ABS OD, Cutoff and Multi standard up to 10 calibrators
Voltage 100-220 V
AC 50 - 60 Hz
Temperature 10° to 40° C
Dimensions 22 x 32 x 16 cms